国产午夜人成在线播放|亚洲人成网站观看在线观看|久久精品国内高清一区二区三区|亚洲精品无码久久毛片

OEM Icon

OEM 無創(chuàng)血壓技術

我們的技術。你的解決方案。

性能最高,尺寸最小的OEM模塊,采用最新硬件平臺,簡單操作可與各種醫(yī)療設備系統(tǒng)集成。 1)在各種的臨床應用環(huán)境中性能優(yōu)異 2)具備心電門控抗運動功能 3)符合醫(yī)療設備第三版及Rohs要求 4) TTL通訊串口 5)小體積,扁平結構便于集成

Adv MX Mini模塊

技術規(guī)格

Physical Dimensions

SPECIFICATIONS VALUES
Width 1.0 in [25.4 mm]
Length 2.75 in [69.9 mm]
Height 0.72 in [18.3 mm]
Weight 1.25 oz. [35.4 g]

Measurement

SPECIFICATIONS VALUES
Technique Oscillometric
NIBP Accuracy Meets ANSI/AAMI SP10, EN 1060-4, ISO 81060-2
Patient Applications Adult, Pediatric, Neonatal
Systolic Range
  • Adult: 30-255 mmHg
  • Neonatal: 30-135 mmHg
Diastolic Range
  • Adult: 15-220 mmHg
  • Neonatal: 15-110 mmHg
Pulse Rate Range
  • Adult: 30-240 BPM (Beats per min)
  • Neonatal: 40-240 BPM
Pulse Rate Accuracy ±2% or ±3 BPM, whichever is greater
Transducer Accuracy ± 3 mmHg over full range in operating conditions
Operating Modes Manual
Operating Conditions 0?C to 55?C, 15% to 90%, non-condensing humidity
Storage Conditions -20?C to +65?C, up to 90% RH non-condensing

Electrical

SPECIFICATIONS VALUES
Operating Voltage 5 VDC ± 5%, 6 VDC ± 5%
Power Consumption Using 5 VDC Nominal
  • Sleep: 50 mW
  • Idle: 150 mW
  • Measurement: 1.0 W
  • Inflation: 3.6 W (High)
Communication Protocols TTL (5V)
Safety & Regulatory Standards IEC 60601-1, IEC/EN80601-2-30, EN60601-1-2:2007, ISO 81060-2:2009

Advantage Technologies

SPECIFICATIONS VALUES
RMT-MAX (Maximum Performance R-wave Motion Tolerance) Yes
TMT-MAX (Maximum Performance Transport Motion Tolerance) Yes
VET-S (Standard Veterinary Monitoring) Yes

Customer Support

SPECIFICATIONS VALUES
Standard Support Yes
Custom Support 10 hours*
Custom Design Additional Charge

Warranty

SPECIFICATIONS VALUES
Time 1 year
Cycles 30000